Double layer custom constant temperature and humidity chamber
Temperature change rates from of 3, 5, 10, 15 or 18K/min.
Advanced Reliability
Lab Companion T/C series has a large tolerance for heat loads generated by the specimen, and achieves improved temperature change rate in an expanded temperature and humidity range.
The new N-instrumentation allows AR series chambers to support an easy-to-use networking system.
-
Two product types
-
-
Rapid-Rate Temperature Cycle Type
-
6K/min per IEC 60068-3-5 (meeting the performance requirement of 5K/min. such as IEC 60068-2-14Nb.) NEW!
-
Standard Type
The product lineup consists of 28 models that can be selected in accordance with temperature & humidity range, inside capacity and temperature change rate.
Test Standard Conformance
IEC 60068-2-1 : Cold
IEC 60068-2-2 : Dry heat
IEC 60068-2-14 Nb : Change of temperature with specified rate of change
IEC 60068-2-30 : Damp heat, cyclic (Rapid-Rate Temperature Cycle Type)
IEC 60068-2-38 : Composite temperature/ humidity cyclic test
IEC 60068-2-78 : Damp heat, steady state
ISO 16750-4(5.3) : Road vehicles, Temperature cycling tests
-
Wide temperature range up to 200°C
-
High temperature testing up to +200°C is possible (option).
-
International safety standards
-
Conforms to ISO12100 and IEC60204, IEC61000 and EN50581 to CE marking requirements based on EU directives: Low voltage directive, EMC directive, machinery directive, pressure equipment directive, and RoHS directive.
-
Remote monitoring and control thanks to a web application
-
No software required, an innovative web browser application allows to monitor and control the chamber with a computer via the chamber standard Ethernet/LAN port. Mobile PC, tablet, and smart phone with multiple chamber connections possible by wireless connection.
-
Email alert
-
When an alarm is triggered, an e-mail is sent to the registered PC or mobile address. A notification can also be sent at the time of test completion.
-
Copy your test programs, output data on USB memory
Without the need of a PC, copy and transfer your test profiles from one chamber to another. You can also download temperature and humidity data from the chamber for editing on your PC.
Rapid-Rate Temperature Cycle Type New!
-
Temperature & humidity range
-
Minimum temp.: -70°C
Maximum temp.: +180°C
Humid.: 10 to 98%rh
Temperature change rate
Capacity
|
270L
|
480L
|
600L
|
800L
|
1000L
|
Temp.range
|
-45?+155°C
|
Heat up
Pull down
|
10K/min.
10K/min.
|
18K/min.
18K/min.
|
10K/min.
10K/min.
|
15K/min.
15K/min.
|
6K/min.
6K/min.
|
10K/min.
10K/min.
|
15K/min.
15K/min.
|
6K/min.
6K/min.
|
* C: temperature & humidity, T: temperature only
-
Totally frost-free
-
Frosting will not appear on any part of the unit despite the temperature & humidity control range of range from 10 to 98%rh. Eradicating the need to remove frosting provides stable and continuous operations.
-
Big tolerance for heat load
-
Temperature only: up to 9000W
Temp. & humid.: up to 550W
Allowable heat load is different depends on models and operation conditions. Please refer to our brochure for the details.
-
Energy-saving with Dual PID Control
-
Dual PID Control (Proportional-Integral-Differential: control that enables the segmentation of refrigeration capacity) mostly controls only the refrigerator at 0°C or lower, thereby saving energy.
-
Specimen temperature control (optional)
-
Attaching a temperature measurement sensor to the specimen enables the temperature of the specimen to be monitored and controlled.
-
Easy access to specimen
-
Equipped with cable ports on both right & left sides as standard.
Door notch port is also available to put cables through the door (option).
-
N-instrumentation equipped with a color LCD touch panel
-
Arge screen with LED backlight is clearer and provides faster display speed.
-
-70 to +180°C (10 to 98%rh)
-
Model
|
Interior
Volume
|
Interior
Dimensions
(mm)
|
Exterior
Dimensions
(mm)
|
Performance
|
Temp. & Humid.
|
Temp. Only
|
Time to Reach Extreme Temp.
|
Temp. Change Rate per
IEC 60068 3-5/ 3-6
|
C/270/70/10
|
T/270/70/10
|
270L
|
W580
H750
D615
|
W895
H1950
D2080
|
+20 → +180°C:
20 min.
+20 → -70°C:
20 min.
|
No
Specimen
|
-45 ? +155°C
Heat-up Rate &
Pull-down Rate:
10°C/min.
|
C/270/70/15
|
T/270/70/15
|
+20 → +180°C:
15 min.
+20 → -70°C:
15 min.
|
No
Specimen
|
-45 ? +155°C
Heat-up Rate &
Pull-down Rate:
18°C/min.
|
C/480/70/10
|
T/480/70/10
|
480L
|
W800
H950
D650
|
W1115
H2090
D2605
|
+20 → +180°C:
20 min.
+20 → -70°C:
20 min.
|
No
Specimen
|
-45 ? +155°C
Heat-up Rate &
Pull-down Rate:
10°C/min.
|
C/480/70/15
|
T/480/70/15
|
+20 → +180°C:
15 min.
+20 → -70°C:
15 min.
|
No
Specimen
|
-45 ? +155°C
Heat-up Rate &
Pull-down Rate:
15°C/min.
|
C/800/70/10
|
T/800/70/10
|
800L
|
W1000
H980
D800
|
W1410
H2085
D2815
|
+20 → +180°C:
20 min.
+20 → -70°C:
20 min.
|
No
Specimen
|
-45 ? +155°C
Heat-up Rate &
Pull-down Rate:
10°C/min.
|
C/800/70/15
|
T/800/70/15
|
+20 → +180°C:
15 min.
+20 → -70°C:
15 min.
|
No
Specimen
|
-45 ? +155°C
Heat-up Rate &
Pull-down Rate:
15°C/min.
|
C/600/70/5
|
T/600/70/5
|
600L
|
W800
H950
D800
|
W1115
H1955
D1855
|
+20 → +180°C:
40 min.
+20 → -70°C:
40 min.
|
No
Specimen
|
-45 ? +155°C
Heat-up Rate &
Pull-down Rate:
6°C/min.
|
C/1000/70/5
|
T/1000/70/5
|
1000L
|
W1100
H950
D950
|
W1415
H2000
D2030
|
+20 → +180°C:
40 min.
+20 → -70°C:
40 min.
|
No
Specimen
|
-45 ? +155°C
Heat-up Rate &
Pull-down Rate:
6°C/min.
|