Rapid-Rate Thermal Cycle Chamber
如果您對(duì)該產(chǎn)品感興趣的話,可以
產(chǎn)品名稱: Rapid-Rate Thermal Cycle Chamber
產(chǎn)品型號(hào):
產(chǎn)品展商: Lab Companion
產(chǎn)品文檔: 無相關(guān)文檔
簡(jiǎn)單介紹
Rapid-Rate Thermal Cycle Chamber
Rapid-Rate Thermal Cycle Chamber
的詳細(xì)介紹
Rapid-Rate Thermal Cycle Chamber
High temperature rate of change with uniformity and reproducibility
The Rapid-Rate Thermal Cycle Chamber is ideal for test requiring quick changes 15°C/min ramp rate in specimen temperature and 23°C/min ramp rate for air temperature(temperature heat up average). It covers various applications from JEDEC, IEC standards to screening.
It features new technologies, such as a specimen temperature control function maintaining linear specimen temperature change rates during rapid thermal cycling, and temperature ramp control.
-
Specifications
-
Model
|
Temperature range
|
Inside / Outside dimensions (mm)
|
T/270/70/25
|
-70 to +180°C (-94 to +356°F)
|
W580×H750×D615 / W895×H1950×D2080
|
Compatible Test standards
IEC-60749-25 : Semiconductor devices – Temperature cycling
IEC-60068-2-14 Nb : Environmental testing – Change of temperature
IEC-61147-5 : Liquid crystal and solid-state display devices – Environmental, endurance and mechanical test methods
JESD22-A105-B : Solid State Devices, testing quality and reliability – Power and temperature cycling
SAE-J1211 : Automotive components - Recommended environmental practices for electronic equipment design
IPC-9701 : Performance test methods and qualification requirements for surface mount solder attachments
Uniformity, reproducibility
Wind simulation allows achieving minimum temperature variations of the specimen, for accurate and quick temperature changes. Specimen temperature ramp rate is 15°C/min, air temperature ramp rate is 23°C/min(temperature heat up average).
Greater temperature ramp control
High precision of the specimen temperature control achieved thanks to a sensor for temperature specimen measurement, and the embedded high-speed controller.
In addition:
-
refrigeration capacities increased at low temperatures,
-
minimization of the difference between test area and specimen temperatures,
-
airflow speed uniformity,
lead to excellent temperature ramp control.
Selection of the temperature control mode
Two temperature control modes are available:
-
A specimen temperature control mode meeting ramp rate conditions of JEDEC standard (15°C/min.)
-
An air temperature control mode for temperature cycle tests.
Free access
Cable ports are equipped on both left and right side, to ease specimen access and wiring.
TEST STANDARD (T/270/70/25, COMPATIBILTY)
Test standard
|
Temperature setting
|
Temperature
change rate
|
Soak time
|
Number
of cycles
|
High temperature (°C)
|
Low temperature (°C)
|
IEC 60749-25
(JESD22-A104-D)
|
G
|
+125 (+15, -0)
|
-40 (+0, -10)
|
Specimen temperature,
15°C / min. or less
|
1, 5, 10, 15 min.
|
Not specified
|
I
|
+115 (+15, -0)
|
-40 (+0, -10)
|
J
|
+100 (+15, -0)
|
0 (+0, -10)
|
K
|
+125 (+15, -0)
|
0 (+0, -10)
|
L
|
+110 (+15, -0)
|
-55 (+0, -10)
|
N
|
+80 (+15, -0)
|
-30 (+0, -10)
|
O
|
+125 (+15, -0)
|
-25 (+0, -10)
|
IEC 60068-2-14 Nb
(JIS C 60068-2-14 Nb)
|
+175 ±2
+155 ±2
+125 ±2
+100 ±2
+85 ±2
+70 ±2
+55 ±2
+40 ±2
+30 ±2
|
-65 ±3
-55 ±3
-40 ±3
-25 ±3
-5 ±3
+5 ±3
|
1±0.2°C / min.
3±0.6°C / min.
5±1.0°C / min.
(AVG)
Average for up to fi ve minutes
|
3 hours,
2 hours,
1 hour,
30 min.,
10 min.
3 hours if not specified in product specifi cations
|
2
|
IEC-61747-5
(EIAJ ED-2531A)
|
+100 ±2
+95 ±2
+90 ±2
+85 ±2
+80 ±2
+75 ±2
+70 ±2
+65 ±2
+60 ±2
+55 ±2
+50 ±2
+45 ±2
+40 ±2
+35 ±2
+30 ±2
|
-50 ±3
-45 ±3
-40 ±3
-35 ±3
-30 ±3
-25 ±3
-20 ±3
-15 ±3
-10 ±3
-5 ±3
-0 ±3
|
1±0.2°C / min.
3±0.6°C / min.
5±1.0°C / min.
(AVG)
Average for up to fi ve minutes
|
3 hours,
2 hours,
1 hour,
30 min.,
10 min.
3 hours if not specified in product specifi cations
|
2
|
JESD22-A105-B
|
A
|
+85 (+10, -0)
|
-40 (+0, -10)
|
6.25°C / min.
|
10 min.
|
1000
|
B
|
+125 (+15, -0)
|
-40 (+0, -10)
|
5.5°C / min.
|
IPC-9701
|
TC1
|
100
|
0
|
Specimen temperature,
20°C / min. or less
|
Specimen temperature,
10 min.
|
200
500
1000
3000
6000
|
TC2
|
100
|
-25
|
TC3
|
125
|
-40
|
TC4
|
125
|
-55
|
TC5
|
100
|
-55
|
IPC-TM-650 2.6.6
|
A
|
+125 (+3, -0)
|
-65 (+0, -5)
|
-
|
30 min.
|
5
|
B
|
+85 (+3, -0)
|
-55 (+0, -5)
|
SAE-J1211
|
+85 to +150
|
-40
|
4 to 6°C / min.
|
Low temperature,
4 hours
|
-
|